Multiplex Immunohisto- chemistry(mIHC)

Issue Type

Core Causes

Solutions

Weak/No Specific Signal

1. Antibody inefficiency (expired, mismatched epitope);

 

2. Inadequate antigen retrieval (insufficient epitope exposure);

 

3. Over-blocking or excessive washing;

 

 4. Incompatible fluorochrome/chromogen

1. Validate antibody specificity and optimize dilution ratio (1:50–1:200);

 

2. Select target-adapted retrieval method (citrate buffer/EDTA, 95–100℃ for 15–20min);

 

3. Adjust blocking time (1–2h at room temperature) and washing cycles (3×5min with PBS-T);

 

4. Choose compatible labels (e.g., fluorochromes with non-overlapping emission spectra for IF-based mIHC)

High Non-Specific Background

1. Excessive antibody concentration;

 

2. Inadequate blocking (improper blocking buffer selection);

 

3. Cross-reactivity between primary and secondary antibodies;

 

4. Unoptimized signal amplification steps

1. Reduce antibody concentration (e.g., adjust from 1:100 to 1:200);

 

2. Block with species-matched serum or 5% BSA;

 

3. Set isotype controls and use monospecific primary antibodies plus cross-adsorbed secondary antibodies;

 

4. Optimize amplification steps (e.g., shorten tyramide signal amplification (TSA) incubation time)

Signal Overlap/Cross-Talk

1. Overlapping fluorochrome emission spectra;

 

2. Incomplete antibody stripping (during sequential staining);

 

3. Cross-reactivity of detection reagents

1. Select fluorochromes with non-overlapping emission spectra (e.g., Alexa Fluor 488/594/647);

 

3. Optimize stripping protocol (e.g., treatment with glycine-HCl buffer pH 2.2 for 10–15min);

 

3. Add reagent-specific blocking steps and verify detection system compatibility

Uneven Staining

1. Uneven antibody incubation (reagent pooling);

 

2. Inconsistent tissue section thickness;

 

3. Inhomogeneous antigen retrieval;

 

4. Precipitates in antibodies/reagents

1. Incubate on a shaker (50–100rpm) to ensure full sample coverage by reagents;

 

2. Standardize section thickness (3–5μm);

 

3. Ensure uniform heating during retrieval;

 

4. Centrifuge reagents (10,000rpm for 5min) to remove precipitates before use

Signal Quenching

1. Poor fluorochrome stability;

 

2. Failure to use anti-fade mounting medium;

 

3. Prolonged light exposure;

 

4. Excessively high ambient temperature

1. Choose photostable fluorochromes (e.g., Alexa Fluor, DyLight series);

 

2. Mount sections with anti-fade mounting medium containing DAPI;

 

3. Reduce excitation light exposure time and use fast-capture imaging systems;

 

4. Perform staining and imaging at 4℃ or on ice whenever possible

Tissue Damage

1. Harsh antigen retrieval conditions (overheating/prolonged duration);

 

2. Excessively high detergent concentration (e.g., Triton X-100);

 

3. Mechanical damage during washing/stripping;

 

4. Incompatible fixative

1. Optimize retrieval conditions (lower temperature/shorten duration);

 

2. Reduce detergent concentration (0.1% Triton X-100);

 

3. Wash gently to avoid vigorous shaking;

 

 4. Fix with 4% paraformaldehyde at room temperature for 15–20min, avoiding over-crosslinking fixation

Inconsistent Signal Intensity Between Runs

1. Antibody batch differences;

 

2. Non-uniform reagent concentrations;

 

3. Uncontrolled incubation temperature/duration;

 

4. Inconsistent tissue processing workflows (fixation/storage)

1. Use antibodies from the same batch, aliquot and store at -80℃;

 

2. Standardize reagent preparation protocols;

 

3. Use temperature-controlled incubators and strictly control incubation time;

 

 4. Standardize tissue fixation (immediate fixation after collection) and storage (store paraffin sections at -20℃)


Related Promotional Journal Downloads

Brand/Complete Product Range Series
Brand/Complete Product Range Series

Detailed introduction of the column

Download and Save
Primary antibodies
Primary antibodies

Detailed introduction of the column

Download and Save
Secondary antibodies
Secondary antibodies

Detailed introduction of the column

Download and Save
mIHC/ Multiplex Fluorescence Detection Kits
mIHC/ Multiplex Fluorescence Detection Kits

Detailed introduction of the column

Download and Save

Explore Our Recommended Popular Products

More products

30,000+ high- quality products available online

Primary Antibodies, Secondary Antibodies, mIHC Kits, ELISA Kits, Proteins, Molecular Biology Products,Cell Lines,Reagents ...

Contact Us